X-ray Analyzers











           ▪ Room, cubicle or area in which an x-ray analyser is operated should have at the

               entrance an illuminated sign or a sign combined with a light, which is activated

               only when the X-ray tube is energised to indicate that the X-ray tube is operating.






           ▪ The lights provided to indicate operation of the X-ray tube should be fail-safe or


               provided with a warning to indicate its failure.






           ▪ Each X-ray analyser should be clearly labelled to indicate whether it is an


               enclosed or a partly enclosed unit.