Page 821 - Radiation Safety
P. 821

Neutron Detector






                                      3 He  vs  BF Detectors
                                                              3


                                    ❑ Boron has a lower neutron cross section (3840 barns) compared

                                                   3
                                          with He (5400 barns), therefore boron counters are less sensitive
                                          than their helium counterparts.


                                                                                                                                             3
                                                                                                                             10
                                    ❑ The energy released per reaction is higher in B than He which
                                          enables BF counters to discriminate against gamma pulses.
                                                             3

                                                                                                                                        3
                                    ❑ BF counters function at much higher voltages than He counters
                                              3
                                          (1 - 4 kV). For voltages above 2 kV it is recommended that a guard

                                          ring be used on the anode insulator. Guard rings prevent

                                          electrical leakage across the insulator and reduce noise from the

                                          counter signal.