Neutron Detector
3 He vs BF Detectors
3
❑ Boron has a lower neutron cross section (3840 barns) compared
3
with He (5400 barns), therefore boron counters are less sensitive
than their helium counterparts.
3
10
❑ The energy released per reaction is higher in B than He which
enables BF counters to discriminate against gamma pulses.
3
3
❑ BF counters function at much higher voltages than He counters
3
(1 - 4 kV). For voltages above 2 kV it is recommended that a guard
ring be used on the anode insulator. Guard rings prevent
electrical leakage across the insulator and reduce noise from the
counter signal.