Calibration and Quality Control







                                  • All dosimeter are characterized by the manufacturer at the time of

                                     production.

                                  • It is compulsory that the dosimeter be calibrated periodically as stipulated

                                     by the appropriate authority.


                                  • Purpose of the calibration:

                                         • Film dosimeter and TL dosimeter are not absolute dosimeter and do not

                                            provide direct reading.

                                         • Film sensitivity decreases with storage time due to environmental exposure
                                            therefore recalibration is required for every six months.


                                         • Sensitivity of TLD may decreases with the number of use, it may due to the
                                            mishandling of the TLD material (TLD chip) –surface scratch, deterioration of
                                            teflon (TLD binder) in TLD card. Recalibration is recommended for every 6

                                            months

                                         • to adjust or determine the response or reading of a dosimeter relative to a
                                            standard or to a series of conventionally true values. (pocket dosimeter)